Theodore I. Malinin, MD
Medical Director and Co-Owner
OsteoLife Biomedical I LLC
1951 NW 7th Avenue
Sanibel Suite
Miami , FL 33136
United StatesFEI 3014623896Issued Mar 31, 2022Office of Biological Products Operations – Division I
Source: FDA, published letter. FDA published this warning letter; View Letter opens FDA's copy. To check it, open FDA's record for this establishment, listed under FEI 3014623896.
Violations4
Issue DateMar 31, 2022
Product TypeBiologics
StatusOpen
Letter Details
Firm
OsteoLife Biomedical I LLC
Location
Theodore I. Malinin, MD
Medical Director and Co-Owner
OsteoLife Biomedical I LLC
1951 NW 7th Avenue
Sanibel Suite
Miami , FL 33136
United States
FEI Number
3014623896
Country
United States
Issue Date
March 31, 2022
Issuing Office
Office of Biological Products Operations – Division I
Product Type
Biologics
Status
Open
Summary
OsteoLife Biomedical I LLC, a biologics manufacturer of Freeze-Dried Particulate Bone products, failed to prevent contamination and cross-contamination during processing of human cellular and tissue-based products in violation of 21 CFR 1271.220(a). The company processed multiple batches from single donors on the same day using identical equipment and technicians without performing cleaning between batches, resulting in post-processing cultures that tested positive for microorganisms including Bacillus species on multiple occasions. Despite detecting positive microbiological cultures in some batches, the company distributed other batches from the same donors, creating a significant communicable disease transmission risk.
OsteoLife Biomedical I LLC FDA Warning Letter, March 31, 2022